Continuous Ventilator Correction: Baxter Healthcare Corporation Issues Correction for Life2000 Ventilator due to a nonconforming battery charger, which triggers a battery alarm and renders the ventilator inoperable

Continuous Ventilator Correction: Baxter Healthcare Corporation Issues Correction for Life2000 Ventilator due to a nonconforming battery charger, which triggers a battery alarm and renders the ventilator inoperable

Baxter Healthcare Corporation is correcting the Life2000 Ventilator System due to a nonconforming battery charger, which triggers a battery alarm and renders the ventilator inoperable. This is a result of a failed crimping operation during manufacturing that caused...
Infusion Pump Battery Recall: ICU Medical Removes Some CSB Batteries Intended For Use With Plum 360, A+, and A+3 Infusion Systems Due to Reports of Allegedly Counterfeit, Untested Batteries In Use

Infusion Pump Battery Recall: ICU Medical Removes Some CSB Batteries Intended For Use With Plum 360, A+, and A+3 Infusion Systems Due to Reports of Allegedly Counterfeit, Untested Batteries In Use

ICU Medical is recalling some CSB batteries that are intended for use with Plum 360, Plum A+, and Plum A+3 Infusion Systems after receiving reports of the allegedly counterfeit batteries being used with infusion pumps. These batteries have not been tested or validated...